Abstract:Objective: A specific GC method was developed for the measurement of Borneol in Boyunding .Methods: A HP-INNOWAX capillary column was used,with FID as detector . Results: The calibration curves of Borneol was in good linearity over the ranges of 0.124 ug·u L-12.480ug·u L-1(r=0.9999).The average recoveries were 99.2% with RSD 0.39%.Conclusions: This method is simple,rapid and accurate with good reproducibility,it is suitable for the quality control of Boyunding.